The Innovate™ System - Complete Workflow Overview

Reagent Constitution & Innovate™ System Setup (KIT4000)

  1. Prepare Reagents:
    • Remove reagents from the refrigerator and place them on the lab bench.
    • Select necessary reagents: ATX, Dilutor DA, Cellsolver, Sensilux, Dilutor DS.
    • Reconstitute ATX and Sensilux using their respective diluents.
    • Mixed reagents are good for 5 days when stored in a cooling unit.
  2. Load Reagents & Start System:
    • Swirl reconstituted reagents to homogenize.
    • Pour them into PP-vials on the Innovate system.
    • Turn on the Innovate Instrument and initiate software.
  3. Microwash & Injector Setup:
    • Ensure Microwash is loaded into PP vials.
    • Place injector tips 1 & 3 into Microwash vials.
    • After the wash cycle, transfer injectors 1 & 3 into reagent PP-vials.
  4. ATP Positive Control Preparation:
    • Reconstitute ATP following the same reagent preparation steps.
    • Dispense 500µL into 25 microcentrifuge tubes and store them in the freezer.
  5. Microtiter Plate Preparation & Testing:
    • For daily controls, dispense 50µL of ATP into two separate wells.
    • Once controls pass, prepare the plate with product samples.
    • Run Time: 1 minute per test.
  6. Automated Testing & Review:
    • Use Innovate software to automate plate preparation.
    • 96-well sample run time: 30 minutes.
    • After completion, remove the plate and review results.
  7. End-of-Day System Maintenance:
    • Place injectors 1 & 3 back into Microwash PP-vials.
    • Run the ‘Wash’ protocol to clean the system.