Reagent Constitution & Innovate™ System Setup (KIT4000)
- Prepare Reagents:
- Remove reagents from the refrigerator and place them on the lab bench.
- Select necessary reagents: ATX, Dilutor DA, Cellsolver, Sensilux, Dilutor DS.
- Reconstitute ATX and Sensilux using their respective diluents.
- Mixed reagents are good for 5 days when stored in a cooling unit.
- Load Reagents & Start System:
- Swirl reconstituted reagents to homogenize.
- Pour them into PP-vials on the Innovate system.
- Turn on the Innovate Instrument and initiate software.
- Microwash & Injector Setup:
- Ensure Microwash is loaded into PP vials.
- Place injector tips 1 & 3 into Microwash vials.
- After the wash cycle, transfer injectors 1 & 3 into reagent PP-vials.
- ATP Positive Control Preparation:
- Reconstitute ATP following the same reagent preparation steps.
- Dispense 500µL into 25 microcentrifuge tubes and store them in the freezer.
- Microtiter Plate Preparation & Testing:
- For daily controls, dispense 50µL of ATP into two separate wells.
- Once controls pass, prepare the plate with product samples.
- Run Time: 1 minute per test.
- Automated Testing & Review:
- Use Innovate software to automate plate preparation.
- 96-well sample run time: 30 minutes.
- After completion, remove the plate and review results.
- End-of-Day System Maintenance:
- Place injectors 1 & 3 back into Microwash PP-vials.
- Run the ‘Wash’ protocol to clean the system.