How-To Video
AllerFlow Gluten Surface Residue Test
Welcome to Hygiena®’s instructional video on AllerFlow Gluten, a rapid and convenient environmental gluten residue test for food processing surfaces.
Overview
- Detects gluten from wheat, barley, rye, and oat with no cross-reactivity to soy, rice, or corn.
- Comprises only two components: a sample collection device and a lateral flow cassette.
- Detects as little as 5 µg of gluten, providing results in 10 minutes.
- Shelf life of up to 15 months, stored at room or refrigerated temperature.
Test Procedure
- Sample Collection
- Allow the swab to reach room temperature before use.
- Label the swab tube with the location.
- Swab a 4x4 inch (10x10 cm) area in a crisscross pattern with consistent pressure.
- Do not touch the swab shaft to avoid contamination.
- Insert swab back into the tube, break the snap valve, and release the extraction buffer by squeezing the bulb.
- Shake gently to mix the sample with the liquid.
- Transferring to the Lateral Flow Device
- Label the lateral flow cassette with the test location.
- Pour the liquid sample into the cassette's fill well until full.
Interpreting Results
- Negative Result: A single blue line (C) means gluten is less than 5 µg—the surface is safe for production.
- Positive Result: A red line (T) alongside the blue control line means gluten exceeds 5 µg—clean and retest.
- Invalid Result: No blue control line—repeat the test with a new device.
Final Notes
- Discard used devices in regular trash.
- Contact Hygiena’s 24/7 customer service for any questions or support.
Now, let’s begin the test!